38 Accessories ACCESSORIES Dynamic calibration Dynamic adjustment and calibration Each of our tablet testing devices contains the standard accessories for calibration with weights and gauge blocks The consistency and linearity of the load cells in the testing devices is demonstrable and outstanding Up to 40 kg of weights would have to be used in order to ensure comprehensive calibration This would be impractical and risky We recommend dynamic calibration for a comprehensive calibration process The Dynamic Calibration Package consists of a certified external load cell a software package for the testing device and a display unit for the dynamic calibration During the dynamic calibration process the hardness measuring station approaches the predefined number of measuring points dynamically The software records the measurement results of the testing device and compares them with the reference load cell A calibration protocol can be printed out at the end ADVANTAGES Comprehensive calibration No calibration weights required ACCESSORIES L x W x H mm DYNAMIC CALIBRATION PACKAGE 152 x 90 x 34 MEASURED PARAMETERS HARDNESS For dynamic calibration the load cell is inserted into the hardness test station of the tablet testing system